DETERMINATION OF CHLORIDE IN AQUEOUS SOLUTION WITH FOODLAB
With the analysis of the chlorides, it is possible to quantify the concentration of salt in the liquids for processing or preserving milk by-products (mozzarella, ricotta ...).
With the FoodLab system it is possible to determine the concentration of chlorides in just a few minutes, without any previous treatment of the sample.
Furthermore, the FoodLab analyzer, using previously dispensed, non-returnable (disposable) cuvettes, requires neither specialized personnel nor reagent preparation.
The method's reliability and precision are guaranteed by a calibration system that makes it possible to align the instrument using standard solutions.
The FoodLab system was tested through recovery tests.
MATERIALS AND CORRELATION METHODS
The FoodLab method is based on a colorimetric reaction in which the chloride ions react with the mercuric thiocyanate and with the Fe (II), forming an orange complex whose intensity, measured at 505 nm, is directly proportional to its concentration in the sample. The FoodLab instrument is capable of measuring the intensity of the color, eliminating the interferences due to the turbidity of the liquids.
Recovery tests were carried out on a sample of liquid for preserving mozzarella (aqueous solution): Sodium chloride in various concentrations was added to 5 aliquots of aqueous solution, which were then tested using FoodLab.
Furthermore, 5 repeatability tests were carried out for each sample.
RESULTS AND DISCUSSION
The FoodLab system allows a practically total recovery of chlorides at the various concentrations. The reference values and the tests carried out using the FoodLab system are compared in Tab.1. The values are perfectly comparable and the repeatability of the data is very good, as seen from the standard deviation and from the VC computed for each sample. In Fig. 1, the regression line correlation coefficient (r = 0.997) confirms the optimum agreement of the instrument with the reference method.
Fig.1 - Correlation between FoodLab method and reference method
Tab.1 - Repeatability tests with FoodLab system
Ref
g/dL
0,20
0,40
0,60
0,80
1,00
 
0,19
0,39
0,61
0,84
0,97
 
0,18
0,40
0,63
0,84
0,95
 
0,18
0,41
0,64
0,79
0,98
 
0,19
0,39
0,63
0,80
1,00
MEAN
0,19
0,40
0,62
0,82
0,98
SD
0,01
0,01
0,01
0,02
0,02
CV,%
6,4
2,1
1,8
2,8
2,0
SHORT TABLE
Linearity
Accuracy
Repeat.
R2
Sensitivity
Total
analysis
time
Test/hour
5 g/dL
±5%
CV < 6%
r > 0,99
0,01 g/dL
5 - 8 min
70
 
 
Foodlab¬
» A new concept in quality analysis of food.
Foodlab is a quick, reliable and practical system for chemical tests on olive oil, seed oils, alimentary fats, vegetable purées, milk, cheese and other dairy products.
FoodLab System is produced by Cdr s.r.l.
CDR Cybernetic Systems for medical, veterinary and food analysis
via degli Artigiani, 6 - 50055
Ginestra Fiorentina - FIRENZE
voice +39055.87143.1 - fax +39055.87143.22
IT 04170450482
www.cdr-mediared.com
e-mail: info@cdr-mediared.com

© Copyright Cdr S.r.l. 1990-2007. All rights reserved.
Powered by mediared