DETERMINATION OF CHLORIDES ON CHEESE BY FOODLAB
In the industrial preparation of certain cheeses, such as mozzarella, it is important to check that the quantity of salt added is of an optimum concentration.

In a few minutes, the FoodLab system permits the determination of the concentration of sodium chloride in homogenised cheese. The preparation of the sample is extremely simplified in respect of the official method (FIL-IDF 17 A). The sample is homogenised in a solution of NaOH 0,25 N. It is not necessary to filter it, centrifuge it and nitric acid is not utilised. The reliability and the precision of the method are guaranteed by a calibration system that permits the alignment of the instrument utilising well-known solutions.

The FoodLab system has been tested via recuperation tests.

MATERIALS AND CORRELATION METHODS
The FoodLab method is based upon a colourmetric reaction in which the chloride ions react with the mercury thiocyanate and with Iron ++ forming an orange substance: its intensity, measured at 505 nm, is directly proportional to its concentration in the sample. The FoodLab instrument is capable of measuring the intensity of the colour by eliminating the interferences caused by the turbidity of the solution.

A sample of mozzarella was homogenised in a solution of NaOH 0,25N and analysed by the FoodLab instrument: 4 aliquots of samples were added with a standard solution of sodium chloride of different concentrations and analysed by the FoodLab instrument.

Furthermore, five repeatability tests were carried out on each sample.
RESULTS AND DISCUSSION
The FoodLab system permitted an almost total recuperation of sodium chloride in the various concentrations. The reference values and the tests carried out utilising the FoodLab system are compared in Tab.1. The values are comparable and the repeatability of the data is very good, as can be seen by the standard deviation and by the VC calculated for each sample. In Fig.1, the correlation coefficient of the straight line of regression (r = 0,998) confirms the optimum alignment of the instrument to the reference values.
Fig.1 - Correlation between FoodLab method and reference method
Tab.1 - Repeatability test using the FoodLab system
Ref
g NaCl
0,67
1,1
1,9
2,5
 
0,65
1,15
1,88
2,45
 
0,67
1,1
1,92
2,5
 
0,69
1,12
1,85
2,52
 
0,63
1,08
1,86
2,55
MEAN
0,66
1,12
1,89
2,51
SD
0,02
0,03
0,04
0,04
CV,%
3,63%
2,42%
1,89%
1,52%
SHORT TABLE
Linearity
Accuracy
Repeat.
R2
Sensitivity
Total
analysis
time
Test/hour
3,5 g %
±5%
CV < 5%
r > 0,99
0,2 g %
5 - 8 min
70
 
 
Foodlab¬
»A new concept in quality analysis of food.
Foodlab is a quick, reliable and practical system for chemical tests on olive oil, seed oils, alimentary fats, vegetable purées, milk, cheese and other dairy products.
FoodLab System is produced by Cdr s.r.l.
CDR Cybernetic Systems for medical, veterinary and food analysis
via degli Artigiani, 6 - 50055
Ginestra Fiorentina - FIRENZE
voice +39055.87143.1 - fax +39055.87143.22
IT 04170450482
www.cdr-mediared.com
e-mail: info@cdr-mediared.com

© Copyright Cdr S.r.l. 1990-2007. All rights reserved.
Powered by mediared